粘度計(jì)英文使用說明書

 

This instrument is widely supplied in measuring axunge paint plastic, food, medication, adhesive, cosmetic and so on like fluidic sticky resistance and liquid absolute viscosity.fix and measuring must be done according requirement from instruction book。
 
Main technic spec
 
·Range of measure:10~105mpa.s
·Rotate speed: 6,12,30,60turn/minute
·Error of measure:±5%(Newton liquid)
·Power:220V±10%
         50Hz
 
Fix

1.     take out chassis base
andsupport from packing
boxcircumrotate three
level 。
2.adjust foot into tapping
hole in chassis base(chart 1)
3.main machine fix:
a. take out main machine by unscrewing the screws in chassis
b. unscrew the handle root screw . inset the power line into handle slot(chart2)
c. penetrate the handle into hole under slot(chart2)
d. screw the handle fix knob to screw the handle and keep the main machine level mostly (chart3)
 
 
8.measure the temperature of measured liquid directly , do not use the temperature value of constant temperature bath
9. it is normal when the viscosity changes because of change of speed and length of measure time during measuring. That is not the instrument hitch.
10.operate seriously according the operation manual
 
 

The whole set

1.main machine————————1
2.(1,2,3,4)rotors———— 1
3.protective support—————— 1
4.support——————————  1
5.lever adjustment screw———— 3
6.manual——————————  1
7.certificate ————————— 1
 
Quotiety table
 
     Run/minute
rotor
60 30 12 6
1 1 2 5 10
2 5 10 25 50
3 20 40 100 200
4 100 200 500 1000
 

 

Notice

1.this instrument should be used in 10—30
2.protective support must be fixed when measuring, or the viscosity value will change as the change of diameter of container
3.the rotor connective screw staff should be raised when fix or remove rotor, or the head of shaft will be damaged. Keep the bleb from rotor when rotor is immerged into liquid
4.the instrument should not be placed in oblique or inverse direction. Especially the rotor is fixed.
5.after fixing the rotor, do not turn on machine to turn the rotor without immerging the rotor into liquid
6.remove rotor carefully after finishing work every time.  Clear it extremely, dry is with blower, and put it back on the rotor support in vertically
7.pointer lock staff should be packaged by yellow package cover and rubber hand when carrying the instrument.
 


 
e. remove the rubber hand from pointer lock staff and yellow package cover from main machine inferior part.
alarm: do not turn on machine before removing the above two objects or it will damage the machine
power on, adjust the level adjustment foot(chart4). Make the bleb in the middle of level set(chart5).
 
Operation
 
1.first rightly rotate the protective support into the instrument inferior part.
2.prepare the liquid for measuring, select the column container having diameter not less than 70 mm
3.control the temperature in liquid for measuring correctly according the precision required by measuring
4. select the right rotor number by estimating the approximate range of viscosity of liquid or measuring, and rotate left the rotor into the connective screw staff(chart4),then rotate the up and down knob(chart5),make the instrument down slowly, immerging into measured liquid gradually, until the rotor liquid level balances the liquid level(chart4).put down the pointer lock staff, power on, relax.

the pointer lock staff,
to make the rotor
circumrotate in the
liquid. The speed is
20-30 run, until the
pointer stay at one
position . put down
the pointer lock staff to
make the reading unchangeable,
then turn off the motor to
make the pointer stay
in the range of reading
window,then get the
reading directly. If the
pointer do not stay in
the frame after turning
off motor, go on to put
down the pointer lock
staff. Repeat the action 
of turning on and off to
make the pointer in the
frame of window. If
selecting 6 and 12 turn /minute,
get the reading directly
without locking the pointer.5.the best is that the read pointer stay 30—90 division range, or change the rotor number or run speed to get the requirement 6.when you can not predict the viscosity value of liquid for measuring , the rotor and turn speed knob should be selected by turning from little value to big value, and from slow to fast, to measure the viscosity value.
 
 
Quotiety  and error correcting
1.the absolute viscosity value of liquid measured can be get by multiplying the reading pointed on dial when measuring and quotiety in quotiety table
that is:
η(mpa.s=k×α
k= quotiety
α=pointer reading
η=viscosity value
2.frequency error correcting
                                     normal frequency
actual viscosity value= indicated value ×—————————
                              actual frequency
 
 

 

Range stable and quotiety table

range stable
rotor Turn/minute
60 30 12 6
1 100 200 500 1000
2 500 1000 2500 5000
3 2000 4000 10000 20000
4 10000 20000 50000 100000
 
 
 

 


[上一篇:烏氏粘度計(jì)原理及其應(yīng)用 ]    [下一篇:美國 BROOKFIELD博勒飛粘度計(jì)使用說明書 ]

使用方法

常見問題

瀏覽更多同類產(chǎn)品

STM-IVB斯托默粘度計(jì)
STM-IVB斯托默粘度計(jì)
 NDJ-1E 旋轉(zhuǎn)粘度計(jì)
NDJ-1E 旋轉(zhuǎn)粘度計(jì)
NDJ-1D 布氏旋轉(zhuǎn)粘度計(jì)
NDJ-1D 布氏旋轉(zhuǎn)粘度計(jì)
烏氏粘度計(jì)
烏氏粘度計(jì)
主站蜘蛛池模板: 北条麻妃一区二区三区av高清| 亚洲综合av在线在线播放| 国产av无码国产av毛片| 亚洲av性色精品国产小电影 | 青草青草视频2免费观看| а天堂中文最新一区二区三区| 大胆欧美熟妇xx| 久久婷婷五月综合色精品| 欧美性猛交xxxx黑人猛交| 少妇久久久久久被弄到高潮| 成全动漫视频观看免费下载| 四川少妇被弄到高潮| 妺七av导航福利| 一个人看的www视频免费观看| 伊人久久大香线蕉无码| 国产免费无遮挡吸乳视频在线观看| 内射人妻无套中出无码| 性一交一乱一伦| 4hu亚洲人成人无码网www电影首页 | 国产精品爽爽va在线观看网站| 亚洲中文字幕久久精品无码a| aaaaa少妇高潮大片| 色狠狠色噜噜av天堂一区| 99久久国产综合精品麻豆 | 成 人 黄 色 网 站 视 频 色| 无码精品视频一区二区三区| 国产青榴视频在线观看| 女同性黄网aaaaa片| 超碰97资源站| 亚洲 自拍 另类小说综合图区| 性色av免费网站| 天堂在线资源中文在线8| 一二三四视频社区在线| 亚洲精品国产精品乱码视色| 樱桃视频影院在线播放| 亚洲av中文无码乱人伦在线观看| 小辣椒福利视频精品导航| 国产99久久精品一区二区| 欧美日韩国产码高清综合人成| 成人精品一区二区三区中文字幕| 亚洲国产av一区二区三区 |